Conceptual design of the gamma-to-electron magnetic spectrometer for the National Ignition Facility
The Gamma-to-Electron Magnetic Spectrometer (GEMS) diagnostic is designed to measure the prompt γ-ray energy spectrum during high yield deuterium-tritium (DT) implosions at the National Ignition Facility (NIF). The prompt γ-ray spectrum will provide "burn-averaged" observables, including t...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2014-11, Vol.85 (11), p.11E122-11E122 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Gamma-to-Electron Magnetic Spectrometer (GEMS) diagnostic is designed to measure the prompt γ-ray energy spectrum during high yield deuterium-tritium (DT) implosions at the National Ignition Facility (NIF). The prompt γ-ray spectrum will provide "burn-averaged" observables, including total DT fusion yield, total areal density (ρR), ablator ρR, and fuel ρR. These burn-averaged observables are unique because they are essentially averaged over 4π, providing a global reference for the line-of-sight-specific measurements typical of x-ray and neutron diagnostics. The GEMS conceptual design meets the physics-based requirements: ΔE/E = 3%-5% can be achieved in the range of 2-25 MeV γ-ray energy. Minimum DT neutron yields required for 15% measurement uncertainty at low-resolution mode are: 5 × 10(14) DT-n for ablator ρR (at 0.2 g/cm(2)); 2 × 10(15) DT-n for total DT yield (at 4.2 × 10(-5) γ/n); and 1 × 10(16) DT-n for fuel ρR (at 1 g/cm(2)). |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.4892900 |