One-dimensional stitching interferometry assisted by a triple-beam interferometer

In this work, we proposed for stitching interferometry to use a triple-beam interferometer to measure both the distance and the tilt for all sub-apertures before the stitching process. The relative piston between two neighboring sub-apertures is then calculated by using the data in the overlapping a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optics express 2017-04, Vol.25 (8), p.9393-9405
Hauptverfasser: Xue, Junpeng, Huang, Lei, Gao, Bo, Kaznatcheev, Konstantine, Idir, Mourad
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this work, we proposed for stitching interferometry to use a triple-beam interferometer to measure both the distance and the tilt for all sub-apertures before the stitching process. The relative piston between two neighboring sub-apertures is then calculated by using the data in the overlapping area. Comparisons are made between our method, and the classical least-squares principle stitching method. Our method can improve the accuracy and repeatability of the classical stitching method when a large number of sub-aperture topographies are taken into account. Our simulations and experiments on flat and spherical mirrors indicate that our proposed method can decrease the influence of the interferometer error from the stitched result. The comparison of stitching system with Fizeau interferometry data is about 2 nm root mean squares and the repeatability is within ± 2.5 nm peak to valley.
ISSN:1094-4087
1094-4087
DOI:10.1364/OE.25.009393