Investigation of formaldehyde and acetaldehyde sampling rate and ozone interference for passive deployment of Waters Sep-Pak XPoSure samplers

This study investigated formaldehyde and acetaldehyde passive sampling rates and ozone interference for the DNPH-based Waters Sep-Pak XPoSure sampler. Previous studies have shown that ozone interferes with active sampling by this cartridge. Our study included one laboratory and six field experiments...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Atmospheric environment (1994) 2013-12, Vol.80, p.184-189
Hauptverfasser: Mullen, Nasim A., Russell, Marion L., Lunden, Melissa M., Singer, Brett C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study investigated formaldehyde and acetaldehyde passive sampling rates and ozone interference for the DNPH-based Waters Sep-Pak XPoSure sampler. Previous studies have shown that ozone interferes with active sampling by this cartridge. Our study included one laboratory and six field experiments conducted in Northern California homes. Passive sampling rates of 1.10 ± 0.09 and 0.86 ± 0.10 mL min−1 determined for formaldehyde and acetaldehyde are lower than previously reported. In a controlled laboratory experiment there were small, statistically insignificant impacts of subsequent ozone exposure on formaldehyde and acetaldehyde mass passively collected on the samplers. This sampler is inexpensive, easy to deploy and to transport by mail, and has a high sampling capacity when used passively; it is suitable for a wide-range of monitoring applications. However, the passive sampling rate remains in question given the internally consistent, but different results obtained in our study and the previous study.
ISSN:1352-2310
1873-2844
DOI:10.1016/j.atmosenv.2013.07.074