이색법을 위한 저가형 흑체복사 교정장치에 관한 연구
The two-color method is known as a relatively easy and reliable method for measuring the temperature of a flame. This method has the advantage that it measures the temperature of a flame without measuring the value of its emissivity. However, to measure temperature using the two-color method, a blac...
Gespeichert in:
Veröffentlicht in: | 한국생산제조학회지 2020, 29(6), , pp.528-533 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | kor |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The two-color method is known as a relatively easy and reliable method for measuring the temperature of a flame. This method has the advantage that it measures the temperature of a flame without measuring the value of its emissivity. However, to measure temperature using the two-color method, a blackbody radiation correction must be performed. In general, radiation correction devices that radiate the blackbody temperature spectrum are quite expensive. Therefore, in this study, blackbody radiation calibration was performed using a tungsten lamp, which has a known emissivity for various temperature ranges. The experimental results clearly indicate that the proposed tungsten lamp configurations can be alternatively applied to calibrate the two-color method as a blackbody source. KCI Citation Count: 0 |
---|---|
ISSN: | 2508-5093 2508-5107 |
DOI: | 10.7735/ksmte.2020.29.6.528 |