Performance Characteristic of the β-Human Chorionic Gonadotropin Assay on the Access 2 for Its Use in Doping Control
The objective of this study was to validate performance characteristics of the Access 2 (Beckman coulter) system for hCG assays for use as a confirmation test for doping control. The Access 2 assay was linear up to 500 IU/L. The correlation coefficient was higher than 0.999, and the sensitivity of t...
Gespeichert in:
Veröffentlicht in: | Bulletin of the Korean Chemical Society 2010, 31(10), , pp.2889-2892 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The objective of this study was to validate performance characteristics of the Access 2 (Beckman coulter) system for hCG assays for use as a confirmation test for doping control. The Access 2 assay was linear up to 500 IU/L. The correlation coefficient was higher than 0.999, and the sensitivity of the linearity was 0.82. There were no false positive or false negative cases. LOD was 0.59 IU/L. The method was robust when performed by different people.
Repeatability and reproducibility were below 7%. We compared reproducibility and recoveries of Access 2 and Elecsys 2010. Access 2 demonstrated higher reproducibility than Elecsys 2010. Recoveries (accuracy) of Access 2were between 85 and 105%. Recoveries for Elecsys 2010 were between 91 and 104%. KCI Citation Count: 1 |
---|---|