A new connecting technique in partial replantation of a ruptured peritoneal dialysis catheter

Abstract Peritoneal dialysis catheter ruptures have been managed by immediate removal and subsequent reinsertion of the catheter which inevitably entails interruption in peritoneal dialysis and a need for vascular access. A 36-year-old man on continuous ambulatory peritoneal dialysis complaining of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Kidney research and clinical practice 2014, 33(3), , pp.154-156
Hauptverfasser: Kim, Yaeni, Dheerendra, Prashant C, Kim, Yong-Soo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Peritoneal dialysis catheter ruptures have been managed by immediate removal and subsequent reinsertion of the catheter which inevitably entails interruption in peritoneal dialysis and a need for vascular access. A 36-year-old man on continuous ambulatory peritoneal dialysis complaining of dialysate leakage was found to have a small rupture near the outer cuff of the peritoneal dialysis catheter. Rather than employing the traditional method of exchanging the whole catheter, a partial replantation procedure to salvage the still-functioning conduit was performed. Two peritoneal dialysis adaptors were used to connect the end of the remaining old catheter to a new extraperitoneal segment of a new catheter and a piece of a transfer set to connect the adaptors. A novel, yet simple and safe, means of partial peritoneal dialysis catheter replantation when managing catheter injuries is suggested.
ISSN:2211-9132
2211-9140
DOI:10.1016/j.krcp.2014.06.001