Shape and Contact Force Estimation of Inserted Flexible Medical Device

Flexible devices such as colonoscopy tube or catheter being inserted into the human-body make contact with the human tissues at several locations. Excessive contact force may damage the human tissues. To cope with this problem, the first aim of this paper is the shape estimation of the flexible body...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of control, automation, and systems 2020, Automation, and Systems, 18(1), , pp.163-174
Hauptverfasser: Ryu, Hwan-Taek, Woo, Jaehong, So, Byung-Rok, Yi, Byung-Ju
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Flexible devices such as colonoscopy tube or catheter being inserted into the human-body make contact with the human tissues at several locations. Excessive contact force may damage the human tissues. To cope with this problem, the first aim of this paper is the shape estimation of the flexible body inside the human body using only two electromagnetic sensors. For this, the flexible body is modeled as kinematically redundant manipulator with many joints and links. The second aim is to estimate contact forces at a point, line contact, and multiple points. Cosserat-rod theory is employed to solve this problem. The effectiveness of the proposed algorithms is proved through both experiment and simulation for a flexible colonoscopy tube.