Performance Evaluation of the Newly Developed BD Phoenix NMIC-500 Panel Using Clinical Isolates of Gram-Negative Bacilli

The emergence of carbapenem resistance among gram-negative bacilli (GNB), mediated by carbapenemase production, has necessitated the development of a simple and accurate device for detecting minimum inhibitory concentrations (MICs) and resistance mechanisms, especially carbapenemase production. We e...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of laboratory medicine 2019, 39(5), , pp.470-477
Hauptverfasser: Park, Byeol Yi, Mourad, Demiana, Hong, Jun Sung, Yoon, Eun Jeong, Kim, Dokyun, Lee, Hyukmin, Jeong, Seok Hoon
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The emergence of carbapenem resistance among gram-negative bacilli (GNB), mediated by carbapenemase production, has necessitated the development of a simple and accurate device for detecting minimum inhibitory concentrations (MICs) and resistance mechanisms, especially carbapenemase production. We evaluated the performance of the BD Phoenix NMIC-500 panel (BD Diagnostic Systems, Sparks, MD, USA) for antimicrobial susceptibility testing (AST) and carbapenemase-producing organism (CPO) detection. We used 450 non-duplicate clinical GNB isolates from six general hospitals in Korea (409 and 41 glucose non-fermenting bacilli [GNFB] isolates). AST for meropenem, imipenem, ertapenem, ceftazidime, and ceftazidime/avibactam, and CPO detection were performed using the Phoenix NMIC-500 panel. Broth microdilution was used as the reference method for AST. The rates of categorical agreement (CA), essential agreement (EA), minor error (mE), major error (ME), and very major error (VME) were calculated in each antimicrobial. In addition, PCR and sequencing were performed to evaluate the accuracy of CPO detection by the BD Phoenix NMIC-500 panel, and the rate of correct identification was calculated. The CA rates were >90% for all antimicrobials tested with the isolates, except for imipenem (87.2%). The GNFB CA rates ranged from 92.7% to 100% for all antimicrobials. The ME rates were 1.7% for and 0% for GNFB. The panel identified 97.2% (243/250) of the carbapenemase-producing isolates. The BD Phoenix NMIC-500 panel shows promise for AST and CPO detection.
ISSN:2234-3806
2234-3814
DOI:10.3343/alm.2019.39.5.470