고온초전도 선재와 정합회로를 이용한 RF coil 제작에 대한 기초연구
The substantial improvement of the signal-to-noise ratio (SNR) can be achieved with small-size samples or low-field MRI system by high-temperature superconducting(HTS) RF coil. The typical HTS RF coil made of HTS thin film is expensive and is limited the coil geometry to planar surface coil. In this...
Gespeichert in:
Veröffentlicht in: | 한국초전도.저온논문지, 14(1) 2012, 14(1), , pp.44-47 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | kor |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The substantial improvement of the signal-to-noise ratio (SNR) can be achieved with small-size samples or low-field MRI system by high-temperature superconducting(HTS) RF coil. The typical HTS RF coil made of HTS thin film is expensive and is limited the coil geometry to planar surface coil. In this study, commercial Bi-2223 HTS tapes was used as RF coil for a 0.35T permanent MRI system. It has advantages of both much lower cost and easier fabrication over HTS thin film coil. SNR gain of the image obtained from the HTS RF coil over a conventional Cu RF coil at room temperature was about 2.4-fold and 1.9-fold using the spin echo pulse sequence and gradient echo pulse sequence respectively. |
---|---|