Downward bias of conductivity based point-of-care hemoglobin measurement compared with optical methods

Background: Point-of-care (POC) arterial blood gas analysis (ABGA) is widely used for checking hemoglobin (Hb) level. However, there is the tendency of downward bias of conductivity-based POC ABGA Hb measurement compared with optical methods. Authors tried to correct that bias by linear regression e...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anesthesia and pain medicine (Korean society of anesthesiologists) 2018, 13(3), , pp.323-328
Hauptverfasser: Heo, Min Hee, Kim, Jun Hyun, Kim, Kyung Woo, Cho, Ho Jae, Choe, Won Joo, Kim, Kyung-Tae, Kim, Ji Yeon, Lee, Sang-Il, Park, Jang Su, Kim, Jung Won
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background: Point-of-care (POC) arterial blood gas analysis (ABGA) is widely used for checking hemoglobin (Hb) level. However, there is the tendency of downward bias of conductivity-based POC ABGA Hb measurement compared with optical methods. Authors tried to correct that bias by linear regression equation. Methods: We retrospectively collected a total of 86 Hb result pairs during surgeries. Hb measured by the Sysmex XE-2100 in the laboratory was set as the gold standard and was compared with that measured by the GEM Premier 3500. Data were compared using the Bland-Altman analysis, the reliability of transfusion decision was assessed using three-zone error grid. The linear regression analysis was performed to find out the relation between the Hb results of POC ABGA and those of laboratory based test. Results: The bias of the Hb measured between Sysmex XE-2100 and GEM Premier 3500 was −0.9 g/dl (P < 0.001, 95% confidence interval, −1.038 to −0.665 g/dl). The percentage error was 16.4%. According to error grid methodology, zone A, B and C encompassed 89.5%, 10.5% and 0% of data pairs. After adjusting the POC ABGA Hb values, the bias of the Hb measured by two methods was 0 g/dl (P = 0.991). The percentage error was 18.2%. The zone A, B and C encompassed 91.9%, 8.1% and 0% of data pairs. Conclusions: Hb measurements obtained with reference to conductivity via a POC ABGA were significantly lower than those obtained via optical methods. This bias may deserve attention of anesthesiologists when POC ABGA Hb level is used as a transfusion guideline. KCI Citation Count: 0
ISSN:2383-7977
1975-5171
2383-7977
DOI:10.17085/apm.2018.13.3.323