Study on the Development of Two-Stage Centrifugal Blood Pump for Cardiopulmonary Support System
In the cardiopulmonary support system with an ECMO (extracorporeal membrane oxygenation), ahigher pump head is demanded for a blood pump. In order to realize a blood pump with higher pump head, higher anti-hemolysis and thrombosis performances, astudy on the development of unprecedented multistage b...
Gespeichert in:
Veröffentlicht in: | International journal of fluid machinery and systems 2014, 7(4), 25, pp.142-150 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In the cardiopulmonary support system with an ECMO (extracorporeal membrane oxygenation), ahigher pump head is demanded for a blood pump. In order to realize a blood pump with higher pump head, higher anti-hemolysis and thrombosis performances, astudy on the development of unprecedented multistage blood pump was conducted. In consideration of the applicationof the blood pump for pediatric patients, a miniature two-stage centrifugal blood pump with the impeller's diameter of40mm was designed and the performance was examined in experiments and computations. Some useful knowledge for a design of the blood pump with higher anti-hemolysis and thrombosis performances was obtained. |
---|---|
ISSN: | 1882-9554 1882-9554 |
DOI: | 10.5293/IJFMS.2014.7.4.142 |