Feasibility of dual-energy computed tomography in radiation therapy planning

In this study, the noise level, effective atomic number ( Z eff ), accuracy of the computed tomography (CT) number, and the CT number to the relative electron density EDconversion curve were estimated for virtual monochromatic energy and polychromatic energy. These values were compared to the theore...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the Korean Physical Society 2017, 71(12), , pp.1056-1063
Hauptverfasser: Sheen, Heesoon, Shin, Han-Back, Cho, Sungkoo, Cho, Junsang, Han, Youngyih
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study, the noise level, effective atomic number ( Z eff ), accuracy of the computed tomography (CT) number, and the CT number to the relative electron density EDconversion curve were estimated for virtual monochromatic energy and polychromatic energy. These values were compared to the theoretically predicted values to investigate the feasibility of the use of dual-energy CT in routine radiation therapy planning. The accuracies of the parameters were within the range of acceptability. These results can serve as a stepping stone toward the routine use of dual-energy CT in radiotherapy planning.
ISSN:0374-4884
1976-8524
DOI:10.3938/jkps.71.1056