Sigma-Metrics of Electrolyte Tests From a Recently Launched New-Generation Proficiency Testing Program of the Korean Association of Quality Assurance for Clinical Laboratory

Dear Editor, A new-generation proficiency testing (PT) was launched by the Korean Association of Quality Assurance for Clinical Laboratory (KAQACL) in early 2016 [1]. Previously, PT for electrolyte tests included two main peer groups, direct and indirect ion-selective electrode (ISE) groups, which w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of laboratory medicine 2017, 37(4), , pp.339-342
Hauptverfasser: Huh, Hee Jin, Park, Yun Mi, Lee, Seungok, Ihm, Chunhwa, Seo, Soyeon, Lee, Sang Gon, Park, Joonseok, Park, Hae Il
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Dear Editor, A new-generation proficiency testing (PT) was launched by the Korean Association of Quality Assurance for Clinical Laboratory (KAQACL) in early 2016 [1]. Previously, PT for electrolyte tests included two main peer groups, direct and indirect ion-selective electrode (ISE) groups, which were independent of the manufacturer and model. However, the current peer group distinguishes the manufacturer of the test system participating in the PT, although it is not a specific test system. Using the revised KAQACL PT results, we evaluated the overall accuracy of Na+, K+, and Cl- tests by the manufacturer, as well as direct and indirect ISE results of electrolyte analyzers currently used in Korea. KCI Citation Count: 1