Walking Pattern Analysis Using an Acceleration Sensor Device
In this paper, a device to analyze gait pattern was developed by using a 2-axis acceleration sensor attached to the foot. The 1st low-pass filter was adapted to limit the frequency band up to 5 Hz. An algorithm to detect the peak value exceeding the threshold voltage of an X-axis acceleration sensor...
Gespeichert in:
Veröffentlicht in: | Journal of electrical engineering & technology 2017, 12(1), , pp.396-401 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this paper, a device to analyze gait pattern was developed by using a 2-axis acceleration sensor attached to the foot. The 1st low-pass filter was adapted to limit the frequency band up to 5 Hz.
An algorithm to detect the peak value exceeding the threshold voltage of an X-axis acceleration sensor and a Z-axis acceleration sensor was developed and normal and abnormal walking patterns were thus differentiated. Also, MCU and Bluetooth were combined to transfer the data to other MCUs to display on an LCD; the size of the device could then be reduced. The new algorithm and the device allowed the individual walking patterns to be easily measured at a low cost and with less restriction on activities compared to conventional multiple pressure sensors or motion camera system. KCI Citation Count: 1 |
---|---|
ISSN: | 1975-0102 2093-7423 |
DOI: | 10.5370/JEET.2017.12.1.396 |