Analysis of the Dosimetric Characteristics of Normoxic Polymer Gel by Magnetic Resonance Images

A new type of polymer gel that eliminates the oxygen chemically instead of using complicated facilities was recently introduced. This study utilized a polymer gel fabricated in a laboratory as a reliable dosimeter for measuring therapeutic radiation. Studies on the optimal MR scan conditions was con...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the Korean Physical Society 2010, 56(3), , pp.874-879
Hauptverfasser: 이석, 김철용, Dae Sik Yang, Young Je Park, Sam Ju Cho, 심수정, 이상훈, 임상욱, Kwang Hwan Cho, Jin Ho Choi, 이레나
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new type of polymer gel that eliminates the oxygen chemically instead of using complicated facilities was recently introduced. This study utilized a polymer gel fabricated in a laboratory as a reliable dosimeter for measuring therapeutic radiation. Studies on the optimal MR scan conditions was conducted. The reproducibility, the accuracy, and the dose rate dependency of the normoxic polymer gel dosimeter were evaluated. The relative dose uncertainty and the relative dose resolution of the polymer gel were also evaluated by scanning it under optimal MR scan conditions. The R2calculated by using echo time (TE) of 20, 40, 60, 80, and 100 ms versus dose was linear for 12 ~ 20 Gy. The polymer gel composed in this study can be used as a reliable dosimeter if it is scanned under optimal MR scan conditions. KCI Citation Count: 4
ISSN:0374-4884
1976-8524
DOI:10.3938/jkps.56.874