Measurement of Cerenkov Light in a Fiber-optic Radiation Sensor by Using High-energy Photon and Electron Beams
In this study, we used a charge coupled device to measure scintillating and Cerenkov light generated in a scintillating fiber-optic radiation sensor irradiated by high-energy photon and electron beams. The intensities of Cerenkov light are measured and characterized as a function of the incident ang...
Gespeichert in:
Veröffentlicht in: | Journal of the Korean Physical Society 2010, 56(3), , pp.765-769 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this study, we used a charge coupled device to measure scintillating and Cerenkov light generated in a scintillating fiber-optic radiation sensor irradiated by high-energy photon and electron beams. The intensities of Cerenkov light are measured and characterized as a function of the incident angles of the high-energy photon and electron beams from a clinical linear accelerator. To minimize or remove Cerenkov light, we investigated a subtraction method using a dummy optical fiber and a wavelength discrimination method using optical filters. Also, the intensities of Cerenkov light induced by high-energy photon and electron beams are compared. KCI Citation Count: 2 |
---|---|
ISSN: | 0374-4884 1976-8524 |
DOI: | 10.3938/jkps.56.765 |