Comparison of noise power spectrum methodologies in measurements by using megavoltage X-ray energies

The noise power spectrum (NPS) is one of the most general methods for measuring the noise amplitude and the quality of an image acquired from a uniform radiation field. The purpose of this study was to compare different NPS methodologies by using megavoltage X-ray energies. The NPS evaluation method...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the Korean Physical Society 2012, 60(1), , pp.129-136
Hauptverfasser: Min, Jung-Whan, Son, Jin-Hyun, Jeong, Hoi-Woun, Kim, Jung-Min, Seoung, Youl-Hun, Son, Soon-Yong, Kim, Ho-Kyung, Kim, Sang-Young, Lee, Do-Wan, Jung, Jae-Yong, Suh, Tae-Suk, Choe, Bo-Young
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The noise power spectrum (NPS) is one of the most general methods for measuring the noise amplitude and the quality of an image acquired from a uniform radiation field. The purpose of this study was to compare different NPS methodologies by using megavoltage X-ray energies. The NPS evaluation methods in diagnostic radiation were applied to therapy using the International Electrotechnical Commission standard (IEC 62220-1). In order to measure the region of interest (ROI) of the NPS, we used the following five factors: the overlapping impact, the non-overlapping impact, the penumbra, the flatness and different ROI sizes. We used NPS from four different types of detectors, the CR-IP (computed radiography image plate: photo-stimulable phosphor screen), the CR-IP-lead (hexalon lead screen), the CR-IP-back [lanex TM fast back screen: {terbium-doped gadolinium oxysulfide granular phosphor screen (Gd 2 O 2 S:Tb,133 mg/cm 2 )} + 1-mm-thick copper plate and the CR-IP-front (lanex TM fast front screen). A Kodak 2000 RT photo-stimulable phosphor-based computed radiographic (CR) system showed that the normalized noise power spectrum (NNPS) curve gradually decreased, in compliance with increasing spatial resolution. In addition, each detector showed a different reactivity of the NPS to megavoltage. The results of multivariate analysis of variance (MANOVA) test (methods × detectors) revealed significant main effects of the methods [F(1, 4) = 53.543, P = 0.001 and of the detectors [F(1, 4) = 17.556, P = 0.001]. The present study revealed that various factors could be employed to produce megavoltage imaging (MVI) of the NPS and as a baseline standard for NPS control in MVI.
ISSN:0374-4884
1976-8524
DOI:10.3938/jkps.60.129