다수 임플란트 지지 보철물에서 지대주 나사의 풀림 토크값에 대한 연구

Purpose: This study evaluated the detorque values of screws in a multiple implant-supported superstructure using stone casts made with 2 different impression techniques. Material and methods: A fully edentulous mandibular master model and a metal framework directly connected to four implants (Br${\a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Taehan Chikwa Poch'ol Hakhoe chi 2010, 48(04), , pp.280-286
Hauptverfasser: 이주리, 이동환, 황재웅, 최정한, Lee, Ju-Ri, Lee, Dong-Hwan, Hwang, Jae-Woong, Choi, Jung-Han
Format: Artikel
Sprache:kor
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: This study evaluated the detorque values of screws in a multiple implant-supported superstructure using stone casts made with 2 different impression techniques. Material and methods: A fully edentulous mandibular master model and a metal framework directly connected to four implants (Br${\aa}$nemark $System^{(R)}$; Nobel Biocare AB) with a passive fit to each other were fabricated. Six experimental stone casts (Group 1) were made with 6 non-splinted impressions on a master cast and another 6 experimental casts (Group 2) were made with 6 acrylic resin splinted impressions. The detorque values of screws ($TorqTite^{(R)}$ GoldAdapt Abutment Screw; Nobel Biocare AB) were measured twice after the metal framework was fastened onto each experimental stone cast with 20 Ncm torque. Detorque values were analyzed using the mixed model with the fixed effect of screw and reading and the random effect of model for the repeated measured data at a .05 level of ignificance. Results: The mean detorque values were 7.9 Ncm (Group 1) and 8.1 Ncm (Group 2), and the mean of minimum detorque values were 6.1 Ncm (Group 1) and 6.5 Ncm (Group 2). No statistically significant differences between 2 groups were found and no statistically significant differences among 4 screws were found for detorque values. No statistically significant differences between 2 groups were also found for minimum detorque values. Conclusion: In a multiple external hexagon implant-supported prosthesis, no significant differences between 2 groups were found for detorque values and for minimum detorque values. There seems to be no significant differences in screw joint stability between 2 stone cast groups made with 2 different impression techniques. 연구 목적: 본 연구는 다수의 임플란트에 의해 지지되는 보철물을, 두 가지 인상법에 의해 얻은 모형들에 연결한 후, 나사의 풀림 토크값을 측정하여 나사 결합부 안정성에 차이가 있는지를 알아보는데 그 목적이 있다. 연구 재료 및 방법: 네 개의 외부 육각 임플란트 (Br${\aa}$nemark $System^{(R)}$; Nobel Biocare AB)에 직접 연결되는 임플란트 상부 구조물과, 이와 수동적 적합을 갖는 완전 무치악 하악주모형을 만들었다. 이 주모형에서 비연결 인상법으로 6회 인상을 채득하여 6개의 모형 (1군)을 제작하고, 아크릴릭 레진을 이용한 연결 인상법으로 6회 인상을 채득하여 6개의 모형 (2군)을 제작하여, 총 12개의 실험모형을 제작하였다. 각 실험모형에 상부 구조물을 지대주 나사 ($TorqTite^{(R)}$ GoldAdapt Abutment Screw; Nobel Biocare AB)로 20 Ncm의 힘으로 조인 후 각 나사의 풀림 토크값을 총 2회 측정하였다. 두 군 간 나사의 풀림 토크값의 비교를 위한 통계분석은 반복측정을 고려하고 나사와 reading을 고정효과로, 모형을 랜덤효과로 포함하는 혼합모형 (mixed model)을 이용하여 유의수준 .05에서 실시하였다. 결과: 나사의 풀림 토크값의 평균은 7.9 Ncm (1군)과 8.1 Ncm (2군)이었으며, 4개 나사의 풀림 토크값 중 최소값인 최소 풀림 토크값의 평균은 6.1 Ncm (1군)과 6.5 Ncm (2군)이었다. 나사의 풀림 토크값은 두 군 간에 통
ISSN:0301-2875
2005-3789
DOI:10.4047/jkap.2010.48.4.280