Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
For high energy photon beams, solid phantom to water dose conversion factors were calculated by using a Monte Carlo method, and the result were compared with measurements and published data. Based on the absorbed dose to water dosimetry protocol, the conversion factor was theoretically divided into...
Gespeichert in:
Veröffentlicht in: | Journal of the Korean Physical Society 2016, 69(4), , pp.647-652 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | For high energy photon beams, solid phantom to water dose conversion factors were calculated by using a Monte Carlo method, and the result were compared with measurements and published data. Based on the absorbed dose to water dosimetry protocol, the conversion factor was theoretically divided into stopping powers ratios, perturbation factors and ratios of absorbed dose to water and that to solid phantom. Data for a Farmer-type chamber and a solid phantom based on polystyrene which is one of the most common material were applied to calculate the conversion factors for 6 MV and 15 MV photon beams. All measurements were conducted after 10 Gy pre-irradiation and thermal equilibrium had been established with solid slabs in a treatment room. The calculated and the measured conversion factors were in good agreement and could be used to confirm the feasibility of the solid phantom as a substitute for water for high energy photon beam. |
---|---|
ISSN: | 0374-4884 1976-8524 |
DOI: | 10.3938/jkps.69.647 |