Spectroscopic properties of large-volume virtual Frisch-grid CdMnTe detectors
CdMnTe(CMT) is a promising alternative material for use as a room-temperature radiation detector. Frisch-grid detectors have a simple configuration and outstanding spectral performance compared with other single-carrier collection techniques. The energy resolution of large-volume virtual Frisch-grid...
Gespeichert in:
Veröffentlicht in: | Journal of the Korean Physical Society 2015, 66(11), , pp.1761-1765 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | CdMnTe(CMT) is a promising alternative material for use as a room-temperature radiation detector. Frisch-grid detectors have a simple configuration and outstanding spectral performance compared with other single-carrier collection techniques. The energy resolution of large-volume virtual Frisch-grid CMT detectors was tested by using several isotopes such as
57
Co,
22
Na,
133
Ba, and
137
Cs together or separately. Energy resolutions of 6.7% and 2.1% were obtained for 122-keV
57
Co and 662-keV
137
Cs gamma rays, respectively, without using any additional signal processing techniques. Also, a 12-mm-thick CMT detector detected the 511-keV and 1.277-MeV gamma peaks of
22
Na with values of the full width at half maximum (FWHM) of 2.7% and 1.5%, respectively. In addition, multiple low- and high-energy gamma peaks of
133
Ba were well separated. The mobilitylifetime product calculated from the shift of the 662-keV photo-peak vs. bias by using Hecht’s equation was 7 × 10
−3
cm
2
/V. These results show the possibility of using CMT detectors in response to various requirements for gamma-ray detection at room-temperature. |
---|---|
ISSN: | 0374-4884 1976-8524 |
DOI: | 10.3938/jkps.66.1761 |