Design study of hard X-ray tomography system to obtain a spatial resolution of 100 nm
An optic-based X-ray tomography system of a high spatial resolution using a conventional X-ray tube was proposed. The system had several X-ray optics: multilayer mirror for monochromatic X-ray, capillary optic for focusing X-ray onto a sample, and objective zone plate. The X-ray tomography system wa...
Gespeichert in:
Veröffentlicht in: | Current applied physics 2012, 12(1), , pp.134-140 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An optic-based X-ray tomography system of a high spatial resolution using a conventional X-ray tube was proposed. The system had several X-ray optics: multilayer mirror for monochromatic X-ray, capillary optic for focusing X-ray onto a sample, and objective zone plate. The X-ray tomography system was designed for obtaining a spatial resolution of 100 nm. Design parameters for each optic were determined and optimized by ray tracing in considering X-ray intensity and reflectivity. The X-ray tomography system with a spatial resolution of 100 nm will provide a good inspect tool in bio-medical field and semiconductor applications.
► The design of X-ray tomography system which can obtain a spatial resolution of 100 nm by X-ray optics is presented. ► The X-ray tomography system is designed for using a conventional X-ray tube. ► Design parameters for several X-ray optics are determined by ray tracing. |
---|---|
ISSN: | 1567-1739 1878-1675 |
DOI: | 10.1016/j.cap.2011.05.010 |