Low-Thermal Conductivity Suspensions Used in the Isolation of the Salt Pills Aboard the Astro-H Adiabatic Demagnetization Refrigerator

An adiabatic demagnetization refrigerator (ADR) utilizes the magnetocholoric effect in a paramagnetic salt to produce sub-Kelvin temperatures. It is a solid-state device that has no moving parts and does not rely upon a density gradient in a working fluid. This makes it ideal for cooling space-based...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Kimball, Mark O., Shirron, Peter J., Wegel, Donald C., James, Bryan L., Galassi, Nicholas M., Faulkner, Richard L., San Sebastian, Marcelino
Format: Report
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An adiabatic demagnetization refrigerator (ADR) utilizes the magnetocholoric effect in a paramagnetic salt to produce sub-Kelvin temperatures. It is a solid-state device that has no moving parts and does not rely upon a density gradient in a working fluid. This makes it ideal for cooling space-based instruments. ·Typically the salt is enclosed in a cylindrical pill that is suspended within the bore of a magnet. The suspension between the salt pill and magnet must be robust enough to survive a launch yet have a thermal conductance that minimizes heat from the magnet that is mechanically, and thermally, anchored to a stage at a higher temperature. Here we detail such a design that uses Kevlar(Trade Mark) as the supporting media in a system that limits motion of the salt pill axial as well as laterally with respect to the magnet bore.