Immobilization of DNA on Quartz Crystal Microbalance Sensor Modified with Self-Assembled Monolayer of Thiol Derivative

[Abstract]: In this study, we investigated the direct detection of DNA, without pretreatment, using a quartz crystal microbalance (QCM) sensor. This sensor is modified by a self-assembled monolayer of a thiol derivative that has an amino group as the terminal functional group. Contact angle values a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Oleo Science 2020-03, Vol.69 (3), p.271-276
Hauptverfasser: Hirobumi Shibata, Junpei Nemoto, Shota Shiba, Yusuke Yamane, Gota Kawai, Kazuaki Hashimoto
Format: Artikel
Sprache:jpn
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:[Abstract]: In this study, we investigated the direct detection of DNA, without pretreatment, using a quartz crystal microbalance (QCM) sensor. This sensor is modified by a self-assembled monolayer of a thiol derivative that has an amino group as the terminal functional group. Contact angle values and the attenuated total reflectance Fourier transform infrared (ATR/FT-IR) spectra of the QCM sensors after immersion into an ethanol solution of thiol derivatives clearly showed that self-assembled monolayers of the derivatives were formed on the QCM sensors. Although QCM measurements using unmodified and carboxylic group-modified sensors could not detect DNA-Na salt, the sensor modified with amino groups could detect the DNA. This system can be used for the analysis of the interaction between DNA and DNA-binding proteins.
ISSN:1345-8957