Validation Study on Five Cytotoxicity Assays by JSAAE-VII. Details of the MTT Assay

The inter-laboratory validation study on 5 cytotoxicity assays conducted by JSAAE has been described in the preceding articles (see Validation Articles I-VI in this issue). Presented here are precise data and the protocols on the MTT assay with two cell lines, HeLa S3 (SC) and SQ-5. Hand-plotted dos...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Alternatives to Animal Testing and Experimentation 1998, Vol.5 (1/2), p.119-130
Hauptverfasser: Hiroshi Itagaki, Tadao Ohno, Masato Hatao, Chiharu Hattori, Akira Hayasaka, Makoto Hayashi, Hiroshi Hori, Koichi Imai, Ikuo Imazeki, Takuya Ishibashi, Miyako Kakuma, Mayako Kato, Akio Kawakami, Michiyo Kitazawa, Shigefumi Kogiso, Satoru Miyazaki, Minako Mori, Keiya Nakajima, Madoka Nakajima, Masaaki Nakamura, Takashi Omori, Hiroshi Ono, Makoto Ono, Yuko Osanai, Kaoru Saijo, Yuka Sano, Tetsuji Sasaki, Shinichiro Sato, Hiroyasu Shimada, Shoichiro Shimogo, Hideaki Sugawara, Shigeo Sugimoto, Minoru Takizawa, Chieko Tamaki, Noriho Tanaka, Noriko Teramoto, Toshiyuki Tsuchiya, Hayao Ueno, Yuji Ugai, Shogo Wada, Shinobu Wakuri, Xinhai Wang, Ikuo Watanabe, Osamu Yamakita, Masanori Yoshida, Isao Yoshimura, Shiro Yukiyama
Format: Artikel
Sprache:jpn
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The inter-laboratory validation study on 5 cytotoxicity assays conducted by JSAAE has been described in the preceding articles (see Validation Articles I-VI in this issue). Presented here are precise data and the protocols on the MTT assay with two cell lines, HeLa S3 (SC) and SQ-5. Hand-plotted dose-response curves provided unexpected information otherwise difficult to assume from calculated ED50 values and accompanying outputs from the logistic analysis program, LAP-JSAAE. The characteristics of data files from 18 laboratories were visualized together with all the log(ED50) values. Variations on negative controls in the assay revealed technical competence characteristics of each laboratory. Generally low OD590 values have been obtained in the present MTT assay but high inter-laboratory reproducibility was observed with both cell lines. These results will supplement our understanding on the MTT assay carried out in the large-scale inter-laboratory validation.
ISSN:1344-0411