Optimizing volumetric modulated arc therapy prostate planning using an automated Fine-Tuning process through dynamic adjustment of optimization parameters

In radiotherapy treatment planning, optimization is essential for achieving the most favorable plan by adjusting optimization criteria. This study introduced an innovative approach to automatically fine-tune optimization parameters for volumetric modulated arc therapy prostate planning, ensuring all...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:PHYSICS & IMAGING IN RADIATION ONCOLOGY 2024-07, Vol.31
Hauptverfasser: Cavus, Hasan, Rondagh, Thierry, Jankelevitch, Alexandra, Tournel, Koen, Orlandini, Marc, Bulens, Philippe, Delombaerde, Laurence, Geens, Kenny, Crijns, Wouter, Reniers, Brigitte
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In radiotherapy treatment planning, optimization is essential for achieving the most favorable plan by adjusting optimization criteria. This study introduced an innovative approach to automatically fine-tune optimization parameters for volumetric modulated arc therapy prostate planning, ensuring all constraints were met. A knowledge-based planning model was invoked, and the fine-tuning process was applied through an in-house developed script. Among 25 prostate plans, this fine-tuning increased the number of plans meeting all constraints from 10/25 to 22/25, with a reduction in mean monitor units per gray without increasing plan's complexity. This automation improved efficiency by saving time and resources in treatment planning.