A Randomized Crossover Trial Comparing Glucose Control During Moderate-Intensity, High-Intensity, and Resistance Exercise With Hybrid Closed-Loop Insulin Delivery While Profiling Potential Additional Signals in Adults With Type 1 Diabetes
OBJECTIVE: To compare glucose control with hybrid closed-loop (HCL) when challenged by high intensity exercise (HIE), moderate intensity exercise (MIE), and resistance exercise (RE) while profiling counterregulatory hormones, lactate, ketones, and kinetic data in adults with type 1 diabetes. RESEARC...
Gespeichert in:
Veröffentlicht in: | DIABETES CARE 2022-01, Vol.45 (1), p.194-203 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVE: To compare glucose control with hybrid closed-loop (HCL) when challenged by high intensity exercise (HIE), moderate intensity exercise (MIE), and resistance exercise (RE) while profiling counterregulatory hormones, lactate, ketones, and kinetic data in adults with type 1 diabetes. RESEARCH DESIGN AND METHODS: This study was an open-label multisite randomized crossover trial. Adults with type 1 diabetes undertook 40 min of HIE, MIE, and RE in random order while using HCL (Medtronic MiniMed 670G) with a temporary target set 2 h prior to and during exercise and 15 g carbohydrates if pre-exercise glucose was |
---|---|
ISSN: | 0149-5992 |