Hybrid closed-loop insulin delivery versus sensor-augmented pump therapy in children aged 6-12 years: a randomised, controlled, cross-over, non-inferiority trial
BACKGROUND: Time in range (TIR) goals are rarely met in children with type 1 diabetes, except at the cost of increased hypoglycaemia episodes. Our objective was to evaluate the safety and efficiency of the Diabeloop DBL4K (Diabeloop, Grenoble, France) hybrid closed-loop system in prepubescent childr...
Gespeichert in:
Veröffentlicht in: | LANCET DIGITAL HEALTH 2022-03, Vol.4 (3), p.E158-E168 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | BACKGROUND: Time in range (TIR) goals are rarely met in children with type 1 diabetes, except at the cost of increased hypoglycaemia episodes. Our objective was to evaluate the safety and efficiency of the Diabeloop DBL4K (Diabeloop, Grenoble, France) hybrid closed-loop system in prepubescent children. METHODS: We did a multicentre, open-label, randomised, controlled, non-inferiority, two-session crossover study in the paediatric endocrinology departments of three university hospitals in France and Belgium. Eligible participants were aged 6-12 years with type 1 diabetes for at least 1 year, glycated haemoglobin A1C 9% (75 mmol/mol) or less, and insulin pump treatment for at least 3 months. Participants were randomly assigned (1:1) to a closed-loop device or sensor-augmented pump (open loop) therapy. Randomisation was by a permuted block randomisation scheme, using an interactive web-based response system, and was stratified on centre (block size 6). The assessed closed-loop device, the Diabeloop for Kids DBL4K hybrid closed-loop system, is an automated blood glucose regulation system composed of a handset, insulin pump, and continuous glucose monitor. The open-loop system is defined as a sensor-augmented pump therapy composed of the usual insulin pump used by the patient and a continuous glucose monitor. A 72-h in-patient period was followed by a 6-week home phase. After a 1-week washout period, the participants crossed over to the other device. The primary outcome, assessed in the intention-to-treat population, was the mean proportion of time spent in hypoglycaemia (3·9 mmol/L [ |
---|---|
ISSN: | 2589-7500 |