Optimization and validation of a LaBr3(Ce) detector model for use in Monte Carlo simulations

A reliable detector model is needed for Monte Carlo efficiency calibration. A LaBr3(Ce) detector model was optimized and verified using different radioactive sources (241Am,133Ba,137Cs,60Co and152Eu) and geometries (point, extended and surface). PENELOPE and MCNP were used for Monte Carlo simulation...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:APPLIED RADIATION AND ISOTOPES 2021-08, Vol.174
Hauptverfasser: Hasan, Md Moudud, Vidmar, Tim, Rutten, Jos, Verheyen, Leen, Camps, Johan, Huysmans, Marijke
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A reliable detector model is needed for Monte Carlo efficiency calibration. A LaBr3(Ce) detector model was optimized and verified using different radioactive sources (241Am,133Ba,137Cs,60Co and152Eu) and geometries (point, extended and surface). PENELOPE and MCNP were used for Monte Carlo simulations. A good agreement was observed between simulated and experimental full energy peak efficiencies (FEPE) as their mean relative difference was 2.84% ± 1.93% and 2.79% ± 1.99% for PENELOPE and MCNP simulation, respectively. The differences between simulated FEPEs of two Monte Carlo codes were negligible except for low energies (< 100 keV).
ISSN:0969-8043