Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully impla...
Gespeichert in:
Veröffentlicht in: | Plos One 2020-04, Vol.15 (4) |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Limited data is available concerning the safety of active middle ear implants (AMEI) during
Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an
actuator as used in a fully implantable AMEI was determined in temporal bones. Eleven
temporal bones were implanted with the actuator according to the manufacturer's surgical
guidelines. The actuator was coupled on the incus short process as recommended for sensorineural hearing loss. Temporal bones were exposed 10 times to the MRI magnetic field
by entering the MRI suite in a clinically relevant way. Computed Tomography (CT) images
were acquired before and after the experiment to investigate the risk of actuator dislocation.
Based on the electrical impedance of the actuator, the loading of the actuator to the incus
was confirmed. Relative actuator displacement was determined on the CT images by comparing the initial with the final actuator position in 3D space. Impedance curves were analyzed after each exposure to check the loading of the actuator to the ossicles. Analysis of
CT images with a 0.30.6 mm in-plane resolution indicate no actuator displacement. The
maximum detected change in impedance for all actuators was 8.43 Ω at the actuator's resonance frequency. Impedance curves measured when the actuator was retracted from the
short process after the experiment still indicate the presence of a clear resonance peak. No
actuator displacement or dislocation could be detected in the analysis of CT images and the
measured impedance curves. Impedance curves obtained when the actuator was retracted
from the incus short process still show a clear resonance peak, indicating the device is still
functional after the MRI exposures. |
---|---|