Nafion-ethylenediamine이 수식된 유리탄소전극에 의한 시차펄스전압전류법으로 철(II) 이온의 정량

Determination of iron(II) ion with a perfluorinated sulfonated polymer(nafion)-ethylenediamine(en) modified glassy carbon electrode was studied. It was based on the chemical reactivity of an immobilized layer(nafion-en) to yield complex $[Fe(en)_3]^{+2}$. The oxidation peak potential by differential...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the Korean Chemical Society 2003, Vol.47 (2), p.115-120
Hauptverfasser: 김경원(Kyong Wone Kim), 김희철(Hee Cheol Kim), 김성현(Sung-Hyun Kim), 박병호(Byung Ho Park), 김연희(Yeon Hee Kim), 김경남(Kyong Nam Kim), 고영춘(Yong Chun Ko)
Format: Artikel
Sprache:kor
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Determination of iron(II) ion with a perfluorinated sulfonated polymer(nafion)-ethylenediamine(en) modified glassy carbon electrode was studied. It was based on the chemical reactivity of an immobilized layer(nafion-en) to yield complex $[Fe(en)_3]^{+2}$. The oxidation peak potential by differential pulse voltammetry(DPV) was observed at 0.340${\pm}$0.015 V(vs. Ag/AgCl). The linear calibration curve was obtained in iron(II) ion concentration range $5{\times}10^{-6}{\sim}0.2{\times}10^{-3} M(0.28{\sim}11.17\; mg/L)$, and the detection limit(3s) was $1.89{\times}10^{-5}$M(1.056 mg/L). Perfluorinated sulfonated polymer(Nafion)-ethylenediamine(en)이 화학수식된 유리탄소전극으로 Fe(II) 이온의 정량에 대해 연구하였다. Fe(II) 이온의 착화제인 en을 nafion에 고정시켜 유리탄소전극 표면에 수식하면 이 수식전극의 en은 Fe(II) 이온과 $[Fe(en)_3]^{+2}$의 착물을 형성한다. Nafion-en이 화학수식된 유리탄소전극에서 시차펄스전압전류법에 의한 Fe(II) 이온의 산화봉우리전위는 0.340${\pm}$0.015 V(vs. Ag/AgCl), 측정범위는 $5{\times}10^{-6}{\sim}0.2{\times}10^{-3} M(0.28{\sim}11.17 mg/L)$, 검출한계(3s)는 $1.89{\times}10^{-5}$M(1.056 mg/L)이었다.
ISSN:1017-2548
2234-8530