닭의 간조직을 이용한 과산화수소 정량 바이오센서
The response characteristics of a new biosensor, which was developed by co-immobilisation of chicken liver tissue and ferrocene in a carbon paste matrix for the amperometric determination of hydrogen peroxide, was evaluated. In the range of electrode potential observed ( - 0.5∼ + 0.05 V vs Ag/AgCl),...
Gespeichert in:
Veröffentlicht in: | Journal of the Korean Chemical Society 1997, Vol.41 (7), p.343-350 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | kor |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The response characteristics of a new biosensor, which was developed by co-immobilisation of chicken liver tissue and ferrocene in a carbon paste matrix for the amperometric determination of hydrogen peroxide, was evaluated. In the range of electrode potential observed ( - 0.5∼ + 0.05 V vs Ag/AgCl), the bioelectrode showed the response time ($t_{95%}$) as low as 13 sec and the detection limit of $5.1{\times}10^{-5}$ M, and exhibited a good selectivity in the case of the addition of the possible interferents tested. Also it offered a high biocatalytic stability and opened up the possibility for the construction of a mediatorless biosensor. 과산화수소 정량을 위하여 닭의 간조직과 ferrocene을 탄소반죽에 혼입시켜 만든 전류법 바이오센서의 감응 특성이 조사되었다. 또한 조사된 전극 퍼텐샬 범위( - 0.5∼ + 0.05 V vs Ag/AgCl) 내에서 $t_{95%}$=13 sec의 빠른 감응시간과 $5.1{\times}10^{-5}$ M(S/N=3)의 검출한계를 보여 주었으며, 감응에 방해가 된다고 여겨지는 물질들에 대하여 좋은 선택성을 보여 주었다. 생촉매 안정성이 좋았으며, 이동매체를 사용하지 않는 전극의 제조 가능성을 제시하였다. |
---|---|
ISSN: | 1017-2548 2234-8530 |