Does Second-Generation Suspensory Implant Negate Tunnel Widening of First-Generation Implant Following Anterior Cruciate Ligament Reconstruction?
Purpose: Tunnel widening following anterior cruciate ligament (ACL) reconstruction is commonly observed. Graft micromotion is an important contributing factor. Unlike fixed-loop devices that require a turning space, adjustable-loop devices fit the graft snugly in the tunnel. The purpose of this stud...
Gespeichert in:
Veröffentlicht in: | Knee surgery & related research 2018-12, Vol.30 (4), p.341 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | kor |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose: Tunnel widening following anterior cruciate ligament (ACL) reconstruction is commonly observed. Graft micromotion is an important contributing factor. Unlike fixed-loop devices that require a turning space, adjustable-loop devices fit the graft snugly in the tunnel. The purpose of this study is to compare tunnel widening between these devices. Our hypothesis is that the adjustable-loop device will create lesser tunnel widening.
Materials and Methods: Ninety-eight patients underwent ACL reconstruction from January 2013 to December 2014. An adjustable-loop device was used in 54 patients (group 1) and a fixed-loop device was used in 44 patients (group 2). Maximum tunnel widening at 1 year was measured by the L’Insalata’s method. Functional outcome was measured at 2-year follow-up.
Results: The mean widening was 4.37 mm (standard deviation [SD], 2.01) in group 1 and 4.09 mm (SD, 1.98) in group 2 (p=0.511). The average International Knee Documentation Committee score was 78.40 (SD, 9.99) in group 1 and 77.11 (SD, 12.31) in group 2 (p=0.563). The average Tegner- Lysholm score was 87.25 (SD, 3.97) in group 1 and 87.29 in group 2 (SD, 4.36) (p=0.987). There was no significant difference in tunnel widening and functional outcome between the groups.
Conclusions: The adjustable-loop device did not decrease the amount of tunnel widening when compared to the fixed-loop device. There was no significant difference in outcome between the two fixation devices.
Level of Evidence: Level 3, Retrospective Cohort |
---|---|