직선 및 곡선 전극(Straight or Curved Electrode)을 사용하여 달걀 흰자에서 시행한 고주파열응고술에서 온도에 따른 병소의 양상 및 크기의 비교

Backgroud: Understanding the electrophysiology of radiofrequency (RF) lesions and determining the size and shape of RF lesions is important for reducing side effects when applied to patients in a clinical setting. We compared the shape and size of RF thermocoagulation produced by straight and curved...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Korean journal of anesthesiology 2000-08, Vol.39 (2), p.260
Hauptverfasser: 이기헌, Gi Hun Lee, 윤선혜, Sun Hye Yoon, 김현주, Hyun Joo Kim, 정배희, Bae Hee Jung, 임소영, So Young Lim, 신근만, Geun Man Shin, 홍순용, Sun Young Hong
Format: Artikel
Sprache:kor
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Backgroud: Understanding the electrophysiology of radiofrequency (RF) lesions and determining the size and shape of RF lesions is important for reducing side effects when applied to patients in a clinical setting. We compared the shape and size of RF thermocoagulation produced by straight and curved 20-gauge electrodes and considered its application in clinical settings. Methods: The white from a hesh hens egg was warmed to 37℃ and placed in a rectangular glass container. Straight and curved 20-gauge electrodes were immersed. The transparency of the egg white and the glass container made it possible to photograph the changes in size of the RF lesions over time. We applied thermocoagulation for 60 seconds at 70, 80, and 90℃. Photographs were taken at 60 seconds. We measured the maximal size of the lesions. A two-way statistical analysis of variance was performed. Results: The thermocoagulations were started at the junction of the insulated and uninsulated portion of the electrode and did not extended beyond the tip. The thermocoagulation size was 4.2 ± 0.8 at 70℃, 6.1 ± 2.9 at 80℃ and 6.1 ± 1.9 at 90℃ using the 20-gauge, 10 mm active tip, straight electrode and 4.5 ± 1.1 at 70℃, 7.2 ± 1.9 at 80℃ and 7.9 ± 2.7 at 90℃ using the 20-gauge, 10 mm active tip, curved electrode. There was no observable difference in the size of the lesions produced by the straight and curved electrodes. Conclusions: We found that temperature was the more important factor in determining lesion size. When the temperature setting is the same, lesions produced by straight and curved electrode of the same gauge are also same size. Therefore the choice of straight or curved electrode should be made to optimize ease of handling and ensure proper location of the electrode tip.
ISSN:2005-6419