Case Report of Anti-Thrombogenicity
Abstract Background: The Toray Filtryzer™-NF is a new polymethyl methacrylate filter with improved hemocompatibility and reduced impact on platelet counts. Objectives: This suggests that, if necessary, a reduction in anticoagulation may be possible when dialysis is performed with the Toray Filtryzer...
Gespeichert in:
Veröffentlicht in: | Blood purification 2023-06, Vol.52 (Suppl 1), p.95-97 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Background: The Toray Filtryzer™-NF is a new polymethyl methacrylate filter with improved hemocompatibility and reduced impact on platelet counts. Objectives: This suggests that, if necessary, a reduction in anticoagulation may be possible when dialysis is performed with the Toray Filtryzer™-NF. Methods: In the following, we dialyzed 5 hemodialysis patients who had a contraindication to full anticoagulation postoperatively or after renal biopsy with the Filtryzer™-NF. Results: A significant reduction in heparin administration was achieved, and in 1 patient, heparin substitution was completely omitted. Despite the significantly reduced heparin doses, no thrombosis of the system occurred during the hemodialysis. Conclusion: In conclusion, hemodialysis using the Toray Filtryzer™-NF is an effective alternative for patients at significantly increased risk of bleeding. |
---|---|
ISSN: | 0253-5068 1421-9735 1421-9735 |
DOI: | 10.1159/000530850 |