Use of a wire extender during neuroprotected vertebral artery angioplasty and stenting

Background The off-label use of an extender wire during vertebral artery stenting and angioplasty with or with neuroprotection has not been previously reported. Material and Methods Retrospective, single-patient, technical report. Results After monorail balloon angioplasty was performed on a proxima...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of NeuroInterventional Surgery 2010, Vol.2 (3), p.198
Hauptverfasser: Lesley, Walter S, Kumar, Ravi, Rangaswamy, Rajesh
Format: Report
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background The off-label use of an extender wire during vertebral artery stenting and angioplasty with or with neuroprotection has not been previously reported. Material and Methods Retrospective, single-patient, technical report. Results After monorail balloon angioplasty was performed on a proximal left vertebral artery stenosis, the 190 cm long Accunet neuroprotection filter device was not long enough for delivery of an over-the-wire stent. After mating a 145 cm long, 0.014 inch extension wire to the filter device, a balloon-mounted Liberté stent was implanted with good angiographic and clinical results. Conclusion The off-label use of an extender wire permits successful over-the-wire stenting on a monorail neuroprotection device for vertebral artery endosurgery.
ISSN:1759-8478
1759-8486
DOI:10.1136/jnis.2010.002311