Use of a wire extender during neuroprotected vertebral artery angioplasty and stenting
Background The off-label use of an extender wire during vertebral artery stenting and angioplasty with or with neuroprotection has not been previously reported. Material and Methods Retrospective, single-patient, technical report. Results After monorail balloon angioplasty was performed on a proxima...
Gespeichert in:
Veröffentlicht in: | Journal of NeuroInterventional Surgery 2010, Vol.2 (3), p.198 |
---|---|
Hauptverfasser: | , , |
Format: | Report |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background The off-label use of an extender wire during vertebral artery stenting and angioplasty with or with neuroprotection has not been previously reported. Material and Methods Retrospective, single-patient, technical report. Results After monorail balloon angioplasty was performed on a proximal left vertebral artery stenosis, the 190 cm long Accunet neuroprotection filter device was not long enough for delivery of an over-the-wire stent. After mating a 145 cm long, 0.014 inch extension wire to the filter device, a balloon-mounted Liberté stent was implanted with good angiographic and clinical results. Conclusion The off-label use of an extender wire permits successful over-the-wire stenting on a monorail neuroprotection device for vertebral artery endosurgery. |
---|---|
ISSN: | 1759-8478 1759-8486 |
DOI: | 10.1136/jnis.2010.002311 |