Use of the McGrath® videolaryngoscope in the management of difficult and failed tracheal intubation

Difficult laryngoscopy and failed tracheal intubation are associated with complications which can be serious, and on occasion, life-threatening. We report three cases of difficult and failed tracheal intubation using a conventional Macintosh laryngoscope in which tracheal intubation was accomplished...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British Journal of Anaesthesia 2008, Vol.100 (1), p.116-119
Hauptverfasser: Shippey, B., Ray, D., McKeown, D.
Format: Report
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Difficult laryngoscopy and failed tracheal intubation are associated with complications which can be serious, and on occasion, life-threatening. We report three cases of difficult and failed tracheal intubation using a conventional Macintosh laryngoscope in which tracheal intubation was accomplished swiftly and easily using a new design of videolaryngoscope, the McGrath®.