A METHOD FOR COMPARING ENDOTRACHEAL CUFFS

Damage to the trachea produced in dogs by large and small residual volume cuffs during 6 h of IPPV was compared using a specially designed endotracheal tube. The cuffs under evaluation were adjusted to exert similar average pressures on the tracheal wall, so that many of the variables believed respo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 1978-05, Vol.50 (5), p.435-444
Hauptverfasser: HOMI, J., NOTCUTT, W., JONES, J. J., SPARKE, B. R.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Damage to the trachea produced in dogs by large and small residual volume cuffs during 6 h of IPPV was compared using a specially designed endotracheal tube. The cuffs under evaluation were adjusted to exert similar average pressures on the tracheal wall, so that many of the variables believed responsible for tracheal injury were controlled. The true compliance of the cuff was measured with the tube inside and outside the trachea of the aneasthetized dogs. The maximum estimated pressure transmitted to the tracheal wall, derived from thes compliance curves, was found to equal the peak airway pressure in the presence of a small air leak past each cuff. At various tracheal wall pressures there were only very minor differences in tracheal damage between the large and small residual volume cuffs tested.