Effect of sternal electrode gap and belt rotation on the robustness of pulmonary electrical impedance tomography parameters

Objective:Non-adhesive textile electrode belts offer several advantages over adhesive electrodes and are increasingly used in neonatal patients during continuous electrical impedance tomography (EIT) lung monitoring. However, non-adhesive belts may rotate in unsedated patients and discrepancies betw...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physiological measurement 2020-04, Vol.41 (3), p.035003-035003
Hauptverfasser: Sophocleous, L, Waldmann, AD, Becher, T, Kallio, M, Rahtu, M, Miedema, M, Papadouri, T, Karaoli, C, Tingay, D G, Van Kaam, A H, Yerworth, R, Bayford, R, Frerichs, I
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Objective:Non-adhesive textile electrode belts offer several advantages over adhesive electrodes and are increasingly used in neonatal patients during continuous electrical impedance tomography (EIT) lung monitoring. However, non-adhesive belts may rotate in unsedated patients and discrepancies between chest circumference and belt sizes may result in a gap between electrodes near the sternum. This project aimed to determine the effects of belt rotation and sternal electrode gap on commonly used lung EIT parameters. Approach:We developed a simulation framework based on a 3D finite-element model and introduced lung regions with little or no ventilation that could be changed according to a decremental positive end-expiratory pressure (PEEP) trial. Four degrees of sternal gap and belt rotation were simulated and their effect on the EIT parameters silent spaces, centre of ventilation, global inhomogeneity index and overdistension/collapsed lung (OD/CL) analysed. Additionally, seven premature infants were examined to assess the influence of leftward and rightward belt rotations in a clinical setting. Main results:Small violations of the electrode equidistance criterion and rotations of the belts less than one electrode space exert only minor effects on the EIT parameters and do not impede the interpretation. Rotations of two and three electrode spaces induce non-negligible effects that might lead to flawed interpretations. The 'best PEEP' determined with the OD/CL approach was robust and identifiable with all studied sternal gaps and belt rotations. Significance:We revealed an important challenge for neonatal EIT applications related to a wide electrode gap at the sternum and belt rotation, which should be avoided in clinical application.
ISSN:0967-3334
1361-6579
1361-6579
DOI:10.1088/1361-6579/ab7b42