Mirror Velocity Sampling Errors in a Michelson Interferometer

Mirror velocity variations in a continuous mirror drive Michelson interferometer result in erroneous interferogram sampling. The effects of both systematic and random mirror fluctuations are evaluated. Four methods of evaluation provide a means to quantify the importance of mirror velocity errors in...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Instrumentation science & technology 1993-01, Vol.21 (3-4), p.163-194
Hauptverfasser: Combs, Roger J., Knapp, Robert B., Field, Paul E.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Mirror velocity variations in a continuous mirror drive Michelson interferometer result in erroneous interferogram sampling. The effects of both systematic and random mirror fluctuations are evaluated. Four methods of evaluation provide a means to quantify the importance of mirror velocity errors in the Fourier spectrometer's performance as a spectroradiometer. The four methods are time interval signal analysis, sine wave test, optical chopper modulation, and examination of consecutive interferograms. Examination of consecutive interferograms includes the monitoring of a constant blackbody infrared radiation source as well as an interposed ammonia sample cell for wavenumber axis registration. Each method furnishes a known input to the interferometer modulation equation (i.e. 2uv=f) and permits an internal validation of the interferometer mirror velocity evaluation results.