Tetraphenylborate-Selective Coated Wire Electrode Based on Tetraphenylborate-Alkyltriphenylphosphonium Derivatives

A coated wire electrode selectively sensitive to tetraphenylborate ion(TPB − ) based on an ion-pair complex of TPB − ion with alkylriphenylphosphonium derivatives was prepared with poly(vinyl chloride)-membrane plasticized with o-nitrophenyloctyl ether. The electrodes responded with Nernstian slopes...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical letters 1996-04, Vol.29 (6), p.879-892
Hauptverfasser: Satake, Hiromu, Nakata, Kenichi, Kaneshina, Shoji
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A coated wire electrode selectively sensitive to tetraphenylborate ion(TPB − ) based on an ion-pair complex of TPB − ion with alkylriphenylphosphonium derivatives was prepared with poly(vinyl chloride)-membrane plasticized with o-nitrophenyloctyl ether. The electrodes responded with Nernstian slopes (ca. 58 mV/decade) over a concentration range of 2x10 −7 -10 −2 M. A change in pH did not affect the electrode performance within the range of pH 5.5-12. The electrodes exhibited very good selectivity for TPB − ion against other organic and inorganic anions. The reproducibility, stablity, response time and lifetime of the electrode were studied. The electrode was used to determine benzalkonium and zephiramine in a bactericidal solution with sodium tetraphenylborate as a titrant.
ISSN:0003-2719
1532-236X
DOI:10.1080/00032719608001441