Turbidimetric Determination of Potassium by Flow Injection Analysis
A turbidimetric flow injection analytical method was developed for the determination of potassium (K) after precipitation with a 3% sodium tetraphenylboron solution at pH 9. The minimum detection limit is about 1.0 μg K/mL. Calibration graphs of absorbance versus concentration are linear up to 20 μg...
Gespeichert in:
Veröffentlicht in: | Analytical letters 1994-06, Vol.27 (8), p.1625-1636 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A turbidimetric flow injection analytical method was developed for the determination of potassium (K) after precipitation with a 3% sodium tetraphenylboron solution at pH 9. The minimum detection limit is about 1.0 μg K/mL. Calibration graphs of absorbance versus concentration are linear up to 20 μg K/mL with a correlation coefficient of 0.9982. The average sampling rate is about 60 samples per hour. Relative standard deviations (s%) obtained with standard solutions of potassium are ca. 1%. Several samples of plant leaves, bottle mineral waters and serum rehydration solutions were analyzed.
* From the PhD Thesis of J.O.R.T. |
---|---|
ISSN: | 0003-2719 1532-236X |
DOI: | 10.1080/00032719408006394 |