Turbidimetric Determination of Potassium by Flow Injection Analysis

A turbidimetric flow injection analytical method was developed for the determination of potassium (K) after precipitation with a 3% sodium tetraphenylboron solution at pH 9. The minimum detection limit is about 1.0 μg K/mL. Calibration graphs of absorbance versus concentration are linear up to 20 μg...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical letters 1994-06, Vol.27 (8), p.1625-1636
Hauptverfasser: de Oliveira Torres, José Roberto, Tubino, Matthieu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A turbidimetric flow injection analytical method was developed for the determination of potassium (K) after precipitation with a 3% sodium tetraphenylboron solution at pH 9. The minimum detection limit is about 1.0 μg K/mL. Calibration graphs of absorbance versus concentration are linear up to 20 μg K/mL with a correlation coefficient of 0.9982. The average sampling rate is about 60 samples per hour. Relative standard deviations (s%) obtained with standard solutions of potassium are ca. 1%. Several samples of plant leaves, bottle mineral waters and serum rehydration solutions were analyzed. * From the PhD Thesis of J.O.R.T.
ISSN:0003-2719
1532-236X
DOI:10.1080/00032719408006394