Determination of Peroxide Value in Vegetable Oils by an Organic-Phase Enzyme Electrode

An organic-phase biosensor is used for measuring the peroxide value in vegetable oils. Rapid flow injection monitoring of the peroxide value is accomplished by injecting the untreated oil sample directly into a chloroform carrier solution which transports it to the downstream horseradish-peroxidase...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical letters 1994-01, Vol.27 (2), p.299-308
Hauptverfasser: Mannino, Saverio, Cosio, M. S., Wang, Joseph
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An organic-phase biosensor is used for measuring the peroxide value in vegetable oils. Rapid flow injection monitoring of the peroxide value is accomplished by injecting the untreated oil sample directly into a chloroform carrier solution which transports it to the downstream horseradish-peroxidase (HRP) based flow detector. The Eastman AQ polymer is used for entrapping the HRP onto the detector surface. The new scheme offers convenient monitoring of low level peroxide values and high selectivity (accrued from the biospecificity of HRP towards peroxide species and from the low operating potential). The biosensor results are consistent with those obtained by the official iodometric method. Such development should facilitate quality control testing of oil products.
ISSN:0003-2719
1532-236X
DOI:10.1080/00032719408001074