Selenium (IV) Selective Electrode

A selenium ion selective electrode is described in which the active ingredient is A liquid membrane consisting of A saturated solution of 3, 3 ′ -diaminobenzidine in hexane. An analytically useful potential change occurs from 10 −4 to 1010 −3 M selenium(IV) at pH 2. 5. The electrode shows fairly goo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical letters 1974-01, Vol.7 (1), p.33-39
Hauptverfasser: Malone, T. L., Christian, G. D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A selenium ion selective electrode is described in which the active ingredient is A liquid membrane consisting of A saturated solution of 3, 3 ′ -diaminobenzidine in hexane. An analytically useful potential change occurs from 10 −4 to 1010 −3 M selenium(IV) at pH 2. 5. The electrode shows fairly good selectivity for selenium over other ions.
ISSN:0003-2719
1532-236X
DOI:10.1080/00032717408059016