Analysis of Bone-Composite Interfacial Strengths
The strength characteristics of the interface between bone and a bone/porous metal composite were analyzed. The metal was 316L stainless steel. The average tensile strength of bone spicules ingrown to the porous metal was 46.5 MPa. This compares to a control bone sample strength of 89.7 MPa. The dif...
Gespeichert in:
Veröffentlicht in: | Artificial cells, blood substitutes, and immobilization biotechnology blood substitutes, and immobilization biotechnology, 1981, Vol.9 (1), p.47-55 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The strength characteristics of the interface between bone and a bone/porous metal composite were analyzed. The metal was 316L stainless steel. The average tensile strength of bone spicules ingrown to the porous metal was 46.5 MPa. This compares to a control bone sample strength of 89.7 MPa. The difference could be due to bone location and bone immaturity. The overall inter facial tensile strength (9.59 MPa) is necessarily lower since the spicules cover only a fraction of the inter facial cross-section. The overall interfacial strength had a wide range of scatter (s = ±4.11 MPa). This was probably due to inherent variations in bone strength with location. A population of samples using porous metal with average pore diameters under 100 μm exhibited measurably lower tensile strengths. |
---|---|
ISSN: | 1073-1199 0090-5488 1532-4184 |
DOI: | 10.3109/10731198109117600 |