Evaluation of a New Disposable Plate Dialyzer and the Adverse Effect of Recirculation Single Pass Dialysis on Clearance
A new disposable plate dialyzer (PF = Travenol Paraflo 1.0 m2 11.5 μ Cuprophan), is compared with an older device of similar design (GL = Gambro Lundia Nova 1.0 m2 13.5 μ Cuprophan). The ulcrafiltration rate (UF) relative to dialyzer pressure (DP) was greater for GL (3.90 ±. 02 DP ml/min) than for t...
Gespeichert in:
Veröffentlicht in: | Journal of dialysis 1977, Vol.1 (7), p.653-664 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new disposable plate dialyzer (PF = Travenol Paraflo 1.0 m2 11.5 μ Cuprophan), is compared with an older device of similar design (GL = Gambro Lundia Nova 1.0 m2 13.5 μ Cuprophan). The ulcrafiltration rate (UF) relative to dialyzer pressure (DP) was greater for GL (3.90 ±. 02 DP ml/min) than for this lot of PF (2.71 ±. 01 DP ml/min). In vivo clearance of urea and creatinine in single pass for PF was 132 ± 5 ml/min and 106 ± 5 ml/min, respectively, at mean blood rate of 200 ml/min. RSP produced significantly lower urea and creatinine clearance (p |
---|---|
ISSN: | 0886-022X 0362-8558 1525-6049 |
DOI: | 10.3109/08860227709037661 |