Evaluation of a New Disposable Plate Dialyzer and the Adverse Effect of Recirculation Single Pass Dialysis on Clearance

A new disposable plate dialyzer (PF = Travenol Paraflo 1.0 m2 11.5 μ Cuprophan), is compared with an older device of similar design (GL = Gambro Lundia Nova 1.0 m2 13.5 μ Cuprophan). The ulcrafiltration rate (UF) relative to dialyzer pressure (DP) was greater for GL (3.90 ±. 02 DP ml/min) than for t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of dialysis 1977, Vol.1 (7), p.653-664
Hauptverfasser: Bierman, Martin H., Easterling, Ronald E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new disposable plate dialyzer (PF = Travenol Paraflo 1.0 m2 11.5 μ Cuprophan), is compared with an older device of similar design (GL = Gambro Lundia Nova 1.0 m2 13.5 μ Cuprophan). The ulcrafiltration rate (UF) relative to dialyzer pressure (DP) was greater for GL (3.90 ±. 02 DP ml/min) than for this lot of PF (2.71 ±. 01 DP ml/min). In vivo clearance of urea and creatinine in single pass for PF was 132 ± 5 ml/min and 106 ± 5 ml/min, respectively, at mean blood rate of 200 ml/min. RSP produced significantly lower urea and creatinine clearance (p