External fixators for pelvic fractures: Comparison of the stiffness of current systems
We evaluated the stiffness of external fixation (EF) systems with a reproducible, standardized human pelvic replica of aluminum and perspex in which a type C pelvic ring injury was created. 12 EF systems were analyzed in 2 situations that necessarily occur during a walking cycle. Endpoints were defi...
Gespeichert in:
Veröffentlicht in: | Acta orthopaedica 2003-04, Vol.74 (2), p.165-171 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated the stiffness of external fixation (EF) systems with a reproducible, standardized human pelvic replica of aluminum and perspex in which a type C pelvic ring injury was created. 12 EF systems were analyzed in 2 situations that necessarily occur during a walking cycle. Endpoints were defined as 15 mm of dislocation or tolerance of the maximum load in each situation. In the no weightbearing situation, all except 2 fixators failed; in the weightbearing situation, all fixators failed. Single bar systems performed better than frame configurations. Stability provided by any external fixator is low, and in the case of a type C pelvic ring injury, it is insufficient for patient mobilization and weightbearing. Single bar systems provide more stability than frames. |
---|---|
ISSN: | 1745-3674 0001-6470 1745-3682 |
DOI: | 10.1080/00016470310013897 |