Design of Dual-Mode Stimulus Chip With Built-In High Voltage Generator for Biomedical Applications
In this work, a dual-mode stimulus chip with a built-in high voltage generator was proposed to offer a broad-range current or voltage stimulus patterns for biomedical applications. With an on-chip and built-in high voltage generator, this stimulus chip could generate the required high voltage supply...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on biomedical circuits and systems 2020-10, Vol.14 (5), p.961-970 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this work, a dual-mode stimulus chip with a built-in high voltage generator was proposed to offer a broad-range current or voltage stimulus patterns for biomedical applications. With an on-chip and built-in high voltage generator, this stimulus chip could generate the required high voltage supply without additional supply voltage. With a nearly 20 V operating voltage, the overstress and reliability issues of the stimulus circuits were thoroughly considered and carefully addressed in this work. This stimulus system only requires an area of 0.22 mm 2 per single channel and is fully on-chip implemented without any additional external components. The dual-mode stimulus chip was fabricated in a 0.25-μm 2.5V/5V/12V CMOS (complementary metal-oxide-semiconductor) process, which can generate the biphasic current or voltage stimulus pulses. The current level of stimulus is up to 5 mA, and the voltage level of stimulus can be up to 10 V. Moreover, this chip has been successfully applied to stimulate a guinea pig in an animal experiment. The proposed dual-mode stimulus system has been verified in electrical tests and also demonstrated its stimulation function in animal experiments. |
---|---|
ISSN: | 1932-4545 1940-9990 |
DOI: | 10.1109/TBCAS.2020.2999398 |