The hydrogen maser cavity step autotuning: theoretical analysis and experimental results

The theoretical analysis of the hydrogen maser cavity autotuning step system produced by the atom line duality modulation method is performed. The expressions for a random error due to the maser frequency fluctuations are obtained. The experimental data of the frequency measurements of CH1-75 masers...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
1. Verfasser: Logachev, V.A.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The theoretical analysis of the hydrogen maser cavity autotuning step system produced by the atom line duality modulation method is performed. The expressions for a random error due to the maser frequency fluctuations are obtained. The experimental data of the frequency measurements of CH1-75 masers, operating in the cavity autotuning mode, are given. The optimal parameters choice of a cavity servo system has allowed to improve the CH1-75 hydrogen maser frequency stability by 3-4 times and achieve Allan deviation (0.7-1.5)/spl middot/10/sup -15/ per 1 d.