The hydrogen maser cavity step autotuning: theoretical analysis and experimental results
The theoretical analysis of the hydrogen maser cavity autotuning step system produced by the atom line duality modulation method is performed. The expressions for a random error due to the maser frequency fluctuations are obtained. The experimental data of the frequency measurements of CH1-75 masers...
Gespeichert in:
1. Verfasser: | |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The theoretical analysis of the hydrogen maser cavity autotuning step system produced by the atom line duality modulation method is performed. The expressions for a random error due to the maser frequency fluctuations are obtained. The experimental data of the frequency measurements of CH1-75 masers, operating in the cavity autotuning mode, are given. The optimal parameters choice of a cavity servo system has allowed to improve the CH1-75 hydrogen maser frequency stability by 3-4 times and achieve Allan deviation (0.7-1.5)/spl middot/10/sup -15/ per 1 d. |
---|---|
ISSN: | 1075-6787 |