Microsecond plasma opening switch as a bremsstrahlung generator
Experiments on efficiency increase of the X-ray sterilizer RS-20 are reported. Increase in X-ray output for a given wall plug power may be obtained due to growth of both electron energy and electron diode current. In the first experiment to increase the electron beam energy (up to 3 MeV) a coaxial P...
Gespeichert in:
Hauptverfasser: | , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Experiments on efficiency increase of the X-ray sterilizer RS-20 are reported. Increase in X-ray output for a given wall plug power may be obtained due to growth of both electron energy and electron diode current. In the first experiment to increase the electron beam energy (up to 3 MeV) a coaxial POS in applied external magnetic field was utilized. An electron diode with anode-converter was parallel to POS. 25% of total current was transmitted to the diode. A dose per pulse was 0.024 Gy at 1 m from the converter. The second experiment used POS with planar anode in the open circuit regime so the anode was utilized for electron beam conversion into bremsstrahlung. |
---|---|
DOI: | 10.1109/BEAMS.1998.816924 |