Paced ECG analysis in mobile cardiac monitor
The main goal of this work was development of ECG processing algorithm for the real-time detection and displaying of implanted pacemaker artifacts, if they are present in the signal. In case of traditionally used characteristics of ECG acquisition circuits and analog-to-digital conversion pacemaker...
Gespeichert in:
Hauptverfasser: | , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The main goal of this work was development of ECG processing algorithm for the real-time detection and displaying of implanted pacemaker artifacts, if they are present in the signal. In case of traditionally used characteristics of ECG acquisition circuits and analog-to-digital conversion pacemaker artifacts are often scarcely distinguished. Wide frequency band of ECG amplifier (0.2-10000 Hz) and high ADC sampling rate (20000 Hz) were used to resolve this problem. All pacemaker spikes detection procedures were realized fully digitally. The developed algorithm testing with the use of the real ECG recordings set showed the following results of pacemaker spikes detection: sensitivity 86.6 % and positive predictivity 99.6 %. The proposed algorithm was included in the ECG processing software of mobile cardiac monitor CardioQVARK based on iPhone series 5, 5S and 6. |
---|---|
ISSN: | 2325-887X |
DOI: | 10.22489/cinc.2016.190-202 |